OmniStar/ThermoStar顶级气体分析系统 |
OmniStar
特點:
Gas Analysis Applications
Option for Corrosive/Explosive Gas Mixtures:
OmniStar for corrosive and explosive gas monitoring is ideally suited for exhaust monitoring of gas processes, process control during production and service/maintenance of the system. The compact system monitors all key functions including purge gas flow and ensures that the scrubber is functioning properly and removing all contaminents.
Applications
Features
Analysis software Quadstar TM
OmniStar | Corrosive/Explosive Series OmniStar | |||||
mass range [amu] | 1-100 | 1-200 | 1-300 | 1-100 | 1-200 | 1-300 |
Gas inlet system | Single valve inlet with stainless steel capillary | Single valve inlet with stainless steel capillary | Single valve inlet with stainless steel capillary | Single valve inlet with stainless steel capillary | Single valve inlet with stainless steel capillary | Single valve inlet with stainless steel capillary |
Pressure reduction | 2-stage, fraction-free |
2-stage, fraction-free |
2-stage, fraction-free |
2-stage, fraction-free |
2-stage, fraction-free |
2-stage, fraction-free |
capillary length (other lengths on request) | 1000mm | 1000mm | 1000mm | 1000mm | 1000mm | 1000mm |
Purge gas requirement | NO | NO | NO | 2-7 Bar of inert gas (not He) 1000sccm flow | 2-7 Bar of inert gas (not He) 1000sccm flow | 2-7 Bar of inert gas (not He) 1000sccm flow |
Order no. 230 V/50-60Hz |
PT M26 250 | PT M26 253 | PT M26 256 | PT M26 290 | PT M26 293 | PT M26 296 |
Order no. 115 V/50-60Hz |
PT M26 251 | PT M26 254 | PT M26 257 | PT M26 291 | PT M26 294 | PT M26 297 |
Dimensions (WxHxL) Wieght |
280 x420 x710 mm (11 x16.5 x28 in.) 45 kg (99.2 lbs.) |
280 x420 x710 mm (11 x16.5 x28 in.) 45 kg (99.2 lbs.) |
280 x420 x710 mm (11 x16.5 x28 in.) 45 kg (99.2 lbs.) |
280 x420 x710 mm (11 x16.5 x28 in.) 45 kg (99.2 lbs.) |
280 x420 x710 mm (11 x16.5 x28 in.) 45 kg (99.2 lbs.) |
280 x420 x710 mm (11 x16.5 x28 in.) 45 kg (99.2 lbs.) |
ThermoStar?
features include:
Analysis software Quadstar TM
Example of Application
TA of calcium oxaiate. The evolving gases water vapour (18), carbon monoxide (28), and carbon dioxide (44) are shown together with the weight and temperature as a function of time.
Option: Gas Stream Selector GSS 300
Heated gas inlet system for up to 12 gas streams
–sample gases and calibration or zero gases
Options | R | Part # |
Test leak (filled with PFTBA) | T | BG442785-T |
Gas inlet with low gas consumption (10-3sccm) | T | BG448845-T |
Automatic inlet pressure regulation, ISO 16KF conn | O | |
1...1000 mbar | BG444816-T | |
5E 10-3...1000mbar | BG444796-T | |
GSS300 Gas stream selector:- | O | |
with 6 sampling streams 230V | PT444750-T | |
with 12 sampling streams 230V | PT444760-T | |
Membrane inlet for liquid analysis | O | BG444490-T |